KNAUER Preparative Module 6100 for Batch Chromatography

Specifications

Ordering Information

Print Version

Documentation

Overview

 
Assistant Modules for Preparative HPLC

The Preparative Module 6100 is a combination module consisting of up to three single modules: Smartline Pump 100 plus a multiport switching valves located on the front and side of the unit.
 
The single instruments as well as the power supply are located in a single housing which easily integrates into either a Smartline tower or a Prepline system. The operation and control of the individual units is not affected by the integration into the module.
 
Sample Dosing System
The HPLC Pump 100 can be equipped with different pump heads for maximum flow rates of 10 or 50 ml/minute in either stainless steel or, for biocompatible applications, titanium materials.
 
This pump can be used for injection of samples into a preparative system, either directly or via a sample loop, in both cases a switching valve isolates the pump from the system itself.
 
Fraction Collection
Fraction collection is realised using the multiport valve located on the side of the module. Valves with either 6, 12 or 16 outlet ports are available. Using the fraction collection control software option from Knauer permits solvent recycling and fraction collection based on either time (volume), signal level, signal slope or even spectral similarity.
 

 

Specifications

Top
 

The actual overall performance specifications of these systems depend to a large extent on the application itself and the actual system configuration. Flowspek AG offers a validation service for the system which will check all agreed specifications for the component modules and ensures that these are all functioning correctly within their specifications. The scope of any validation is usually agreed with our customers and then controlled validation document created.

 

Smartline 100 Pump

Delivery System

Dual piston pump with main and auxiliary piston

Flow ranges (ml/min)

0.1 - 9.99 ml/min

0.1 - 49.9 ml/min

Pressure

max. 400 bar

max. 150 bar

Flow Rate Accuracy

better then 0.5 % at 1 ml/min RSD (Relative Standard Deviation)

Display

Flow is permanently displayed

System Protection

Analogue setting of Pmax

Control

RS-232 Interface, Remote Start input, Output for error conditions

All specifications may be subject to change for technical reasons.

 

 
Ordering Information
 

Top

A5004V01

Smartline Prep-Modul with Pump 100 (50ml pump head without pressure transducer), 2 valve drives for 6-Port valve and 7-Port-valve, 1/8"

A5004V02

Smartline Prep-Modul with Pump 100 (50ml pump head without pressure transducer), 2 valve drives for 6-Port valve and 7-Port-valve, 1/16"

A5004V03

Smartline Prep-Modul with Pump 100 (50ml pump head without pressure transducer), valve drive for 6-Port valve, valve drive for 13-Port-valve, 1/8"

A5004V04

Smartline Prep-Modul with Pump 100 (50ml pump head without pressure transducer), valve drive 6 with 6-Port valve, valve drive 16 with 17-Port-valve, 1/16"

A5004V06

Smartline Assistant with Pump 100 (10ml pump head without pressure transducer), valve drive 6 with 6-Port 1/8" valve

A5004V07

Smartline Assistant with valve drive 6 with 6-Port 1/8" valve, valve drive for 13-Port-valve, 1/8"

 