Smartline HPLC UV Detector 2500

Specifications

Ordering Information

Print Version

Documentation

Overview

 
Smartline UV HPLC Detector 2500

The Smartline HPLC Detector 2500 is a programmable variable wavelength detector for HPLC and has been developed specially for use in routine analytical work.
 
Programmable Wavelength
The measuring wavelength is programmable and can be modified during a run. In "Stop Flow" mode spectra can also be acquired.
 
GLP Functions
The detector provides GLP functions such as: detailed reports, displaying the total hours of lamp operation, total number of lamp ignitions, and more.
 
Fiber-Optic Options
The use of fiber-optic guides permits the spacial separation of the HPLC Detector 2500 and flow cell, which makes operation with externally thermostatted cells or in hazardous areas possible. Fiber-optic guides are available in standard or customised lengths.
 
Large Selection of Flow Cells
A large range of flow cells, from nano-HPLC cells for flowrates above 100 nL/minute to preparative cells for flowrates up to 10000 ml/minute, offers maximum flexibility and permits use of the HPLC Detector 2500 for practically all HPLC applications.
 
Service Friendly
The entire optical unit slides out of the detector for easy exchange of lamps etc. The housing does not need to be removed thus making system servicing fast and easy.
 
Compact
The HPLC Detector 2500 has been designed to fit directly into the Smartline instrument family as part of a system tower. The modular design of the detector means that it can also be used as a completely separate HPLC module and its small size allows it to be installed practically anywhere.
 

 

Specifications

Top
 

Wavelength range

190 - 740 nm, 2nd order cut-off filter at 370 nm

Light source

Deuterium lamp as standard, optional Tungsten-Halogen lamp

Bandwidth

8 nm

Wavelength accuracy

± 2 nm

Wavelength verification

Holmium oxide filter cell

Baseline drift

< 1.5 x 10-4 AU/h, ASTM E1657-94

Noise

< 1 x 10-5 AU, ASTM E1657-94

Scan function

In "Stop Flow" mode only

Integrator output

±1 V adjustable in 16 ranges

Data acquisition

Analogue via Integrator output

Display

LCD, 2 x 24 characters

Time constant

0.1; 0.2; 0.5; 1.0; 2.0; 5.0: 10.0 seconds

Remote control

Remote via RS-232 interfaces, analogue inputs for autozero, start program and lamp off, analogue outputs for error conditions
(The detector can be controlled via Geminyx Software.)

Local control

Touchpad

GLP

Recording of total running time and number of lamp ignitions.

Features

Programmable in Stand-alone mode

Dimensions

226 x 135 x 390 mm (W x H x D)

Weight

6 kg

All specifications may be subject to change for technical reasons.

 

 
Ordering Information
 

Top

A5140

Smartline UV Detector 2500 without flow cell, with accessories for operation without fiber-optics

A5141

Smartline UV Detector 2500 without flow cell, with accessories for operation with fiber-optics

Analytical Flow Cells

A4061

10 mm path, 1/16’’ connections, 10 µL, stainless steel, max. flow 20, max. pressure 300 bar

A4042

3 mm path, 1/16’’ connections, 2 µL, stainless steel, max. flow 50, max. pressure 300 bar

A4045

3 mm path, 1/16’’ connections, 2 µL, PEEK, max. flow 50, max. pressure 30 bar

Preparative Flow Cells

A4066

0.5 / 1.25 / 2 mm path, 1/8’’ connections, stainless steel, max. flow 1000, max. pressure 200 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4068

0.5 / 1.25 / 2 mm path, 1/4’’ connections, stainless steel, max. flow 10000, max. pressure 200 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4067

0.5 / 1.25 / 2 mm path, 1/8’’ connections, PEEK, max. flow 1000, max. pressure 100 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4069

0.5 mm path, 1/16’’ connections, stainless steel, max. flow 250, max. pressure 200 bar

A4095

0.5 mm path, 1/16’’ connections, PEEK, max. flow 250, max. pressure 100 bar

Analytical flow cells with double fiber optical connection

A4074

10 mm path, 1/16’’ connections, 10 µL, stainless steel, max. flow 20, max. pressure 300 bar

A4044

3 mm path, 1/16’’ connections, 2 µL, stainless steel, max. flow 50, max. pressure 300 bar

A4047

3 mm path, 1/16’’ connections, 2 µL, PEEK, max. flow 50, max. pressure 30 bar

A4077

3 mm path, 1/16’’ connections, 0.7 µL, PEEK, max. flow 10, max. pressure 30 bar

Preparative flow cells with double fiber optical connection

A4078

0.5 / 1.25 / 2 mm path, 1/8’’ connections, stainless steel, max. flow 1000, max. pressure 200 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4079

0.5 / 1.25 / 2 mm path, 1/8’’ connections, PEEK, max. flow 1000, max. pressure 100 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4081

0.5 / 1.25 / 2 mm path, 1/4’’ connections, stainless steel, max. flow 10000, max. pressure 200 bar
(Cell volumes for adjustable cells are: 0.50 mm = 1.7 µL; 1.25 mm = 4.3 µL;2.00 mm = 6.8 µL.)

A4089

0.5 mm path, 1/16’’ connections, stainless steel, max. flow 250, max. pressure 200 bar

A4096

0.5 mm path, 1/16’’ connections, PEEK, max. flow 250, max. pressure 100 bar

U-Z ViewTM Micro Flow Cells (not available with fiber optics)

A4091

8 mm path, 1/16’’ connections, 0.140 µL, fused silica, max. flow 0.1, max. pressure 500 bar

A4092

8 mm path, 280 µm connections, 0.035 µL, fused silica, max. flow 0.01, max. pressure 500 bar

Spares and Accessories

A0895

RS232 cable for serial PC-interface

A4071-1

Deuterium lamp (35W)

A4072

Tungsten-Halogen lamp

A4073

Tungsten-Halogen lamp with base (required for first time installation)

A0911

Fiber-optic connector, length 400mm

A0742

Fiber-optic connector, length 750mm

A0740

Fiber-optic connectors, length 750mm, pack of 2 items

A0743

Fiber-optic connector (individual length to be specified with order)

A0857

Extension Kit for coupling of two fiber-optic connectors

A4126

Test cell, holmium oxide filter for detectors without fiber-optics

A4128

Test cell, holmium oxide filter for detectors with fiber-optics

A1447

1 lens, 1 seal for analytical flow cells

A1448

1 lens, 1 seal for preparative flow cells

A1449

1 lens, 1 seal for flow cells using fiber optics

A1131

Repair kit for analytical flow cells, 10mm

A1475

Repair kit for analytical flow cells, 3mm

A1132

Repair kit for preparative flow cells

 