flowspek ag
HPLC - UV/VIS Spectroscopy 
Fiber Optic and Diode Array Applications 
Software - Fluorescence 
> SPECTROSCOPY > UVIKON XL UV/VIS Spectrophotometer
 
Overview Specifications Ordering Information Print Version Documentation
Overview
UVIKON XLThe UVIKON XL design is based on the UVIKON Simplicity Concept, which represents the first Spectrophotometer Series featuring Advanced Digital Signal Processing Technology.
 
The UVIKON XL uses state-of-the-art photomultiplier detection, in combination with the high-speed grating drive this permits optimised rapid scanning capability.
 
True Symmetrical Double-Beam Optics
The UVIKON optical design features true symmetrical double-beam optics using a single chopper wheel and detector. The speed of rotation of the chopper wheel allows for a measurement every 10 milliseconds on both the sample and reference channels.
 
This means that effectively all measurements are made in real time against the reference thus providing maximum confidence in the measured result.
 
High Energy Throughput
The UVIKON uses a high-efficiency grating designed by Zeiss for maximum energy throughput. This, combined with the specially designed high-speed grating drive motor, results in precise measured data not compromised by noise.
 
Large Sample Compartment
Total and easy access to the large sample compartment (140 mm deep, 332 mm wide and 162 mm high) with 120 mm beam separation provides for ease-of-use and great flexibility.
 
"Quick-Lock" Positioning
The three-point "Quick-Lock" positioning of accessories gives high reproducibility of results and also allows for fast exchange of accessories whenever required.
 
Protected Optics Area
Quartz windows at the entrance and exit points to the sample compartment prevent contamination of the optics area. The complete optical system is sealed against atmospheric contamination by dust and/or volatile sample compounds.
 
Thermally Isolated Lamp Housing
External positioning of the light source bench provides for thermal insulation of the optical bench.
 
Built-in Accessory Control
The UVIKON XL comes complete with a built-in accessory controller for control of the accessories, e.g. automated cell changer.
SpecificationsTop
Wavelength range180 - 900 nm
Wavelength step0.05 - 10 nm
Scan speed5 - 5000 nm/min
Transfer speed7000 nm/min
Spectral bandwidth0.2 - 0.5 - 1 - 2 - 4 nm;
Reduced height: 0.5 - 1 - 2 - 4 nm
Wavelength accuracy
Holmium Oxide Filter
± 0.25 nm
Wavelength precision
Holmium Oxide Filter
± 0.025 nm
Baseline flatness< ± 1 mabs measured over 200-800 nm, 2 nm, 200 nm/min
Stray light
Nal, 220 nm
< 0.015 %T
Linearity
250 nm, 2 nm, 1 sec.
3.3 Abs, (r2 >= 0.999)
Photometric range± 5 Abs
Photometric accuracy± 3 mabs measured at 590 nm, certified 0.85 AU filter, 1 sec
Photometric precision± 0.5 mAbs measured at 590 nm, certified 0.85 AU filter, 1 sec
Noise (RMS)< 40 µ Abs measured at 580 nm
Drift
< 0.1 mAbs/hr measured at 580 nm after warm-up
Response time0.02 - 5 sec
Light sourcesDeuterium and Tungsten-Halogen lamps
MonochromatorHigh energy low stray light diffraction grating with 1300 lines/mm, 175 mm focal length
Dimensions680 x 275 x 565 mm (W x H x D)
Weight35 kg
All specifications may be subject to change for technical reasons.
Ordering InformationTop
99-90283UVIKON XL Spectralphotometer
(Includes LabPower Software Package, Standard Thermostatable Cell Holder compatible with Micro and Ultra Micro Cells, Accessory Controller to allow connection of all UVIKON accessories)
91-00580Double Magnetic stirrer accessory (variable speed)
91-00616Magnetic stirring bars (x5), size: 8 x 3 mm
91-00606Quick-Lock cell holder for cells up to 20 mm pathlength, water thermostatable
91-00610Thermostation fluid connection block
91-00601Quick-Lock long path rectangular cell holder for cells up to 100 mm pathlength
91-00603Quick-Lock long path cylindrical cell holder for cells up to 100 mm pathlength
91-00624Quick-Lock solid sample holder
91-00586Quick-Lock automatic/manual cell changer, water thermostatable. Modes [6+6]-[6+1]-[2(5+1)]
91-00607Temperature sensor for cell holder
91-00608Inert glass miniaturised temperature sensor for cell content
91-91465ThermoPack Peltier thermoregulated bath from 20 to 60°C
91-91637ThermoSystem Peltier thermoregulated 6x6 cell changer 0 to 100°C and 12 position Stirrer
91-91671Peristaltic sipper thermostatable (requires one circulation cell)
H-176703x1 standard quartz flow cell for sipper volume 160 µl, 10 mm pathlength, Z 15 mm
91-91672Integrating Sphere, 200-850 nm, diameter 63,5 mm, Photomultiplier R928
91-91673Non-calibrated Spectralon reflectance standard
91-91274Calibrated Spectralon reflectance standard
91-912621" Diameter cone-type light trap
91-91268Powder and cream cell holder
96-90275Pre-adjusted halogen lamp for UVIKON XS or XL
96-90297Pre-adjusted deuterium lamp for UVIKON XS or XL
Top of Page | Home-PageCopyright © 2002-dato by flowspek ag
www.flowspek.ch | info@flowspek.chEfringerstrasse 32, CH-4057 Basel